HASETRI Calibration Lab provides accredited Calibration for all types of Temperature sensors including SPRT, RTD and Thermocouples with or without indicator, Digital Thermometer, Liquid-in-glass thermometers, Temperature indicator and controllers of baths, ovens and furnaces.

SPRT, PRT, RTD Calibration:

Master Instruments Used: SPRT, Liquid Bath, Dry Block Calibrator & Digital Thermometer readout

Range:

  • -80°C to 100°C with CMC value of ±0.05°C
  • 100°C to 300°C with CMC value of ±0.07°C
  • 300°C to 650°C with CMC value of ±0.10°C

Thermocouple Calibration:

Master Instruments Used: Standard R-Type Thermocouple with CJC Probe, SPRT, Liquid Bath, Dry Block Calibrator & Digital Thermometer readout

Range:

  • -80°C to 100°C with CMC value of ±0.06°C
  • 100°C to 650°C with CMC value of ±0.10°C
  • 650°C to 1200°C with CMC value of ±1.90°C

Liquid in Glass Thermometer:

Master Instruments Used: SPRT, Liquid Bath & Digital Thermometer readout

Range:

  • -40°C to 200°C with CMC value of ±0.05°C
  • 200°C to 300°C with CMC value of ±0.06°C

Oven/Freezer Calibration (9-Point Calibration) :

Master Instruments Used: Calibrated 1/10 DIN Standard RTDs with Multi Channel Data Acquisition System with DAQ Software

Range:

  • -40°C to 250°C with CMC value of ±0.5°C

Relative Humidity Sensor & Indicator Calibration:

Master Instruments Used: RH Calibrator, Rh & Temperature Sensor with digital indicator

Range:

  • 15% to 95% rH at RT with CMC value of ±0.85%